Introduction:
Recent advances to make cardiopulmonary bypass more physiological include the use of kinetic-assisted venous drainage but without a venous reservoir. Despite manipulation of intravascular volume and patient positioning, arterial flow is frequently reduced. Negative venous line pressures can be generated, which may elicit gaseous microemboli. We investigated the influence of venous cannula design on venous return and negative venous line pressures.
Methods:
In a single-centre, single-surgeon, prospective, randomized, double-blind trial, 48 patients undergoing isolated coronary artery, aortic valve or combined coronary artery and aortic valve surgery, with a minimally invasive circuit, were randomized to a conventional two-stage (2S) or three-stage venous cannula (3S), or to a three-stage venous cannula with additional ‘fenestrated’ ridges (F3S). Blood flow, venous line pressures and gaseous microemboli number and size were measured.