Introduction:
The use of volatile agents during cardiopulmonary bypass allows a “single drug anesthesia” and is associated with reduced peak postoperative troponin levels. Connecting the exhaust systems to the oxygenator’s gas outlet port is mandatory and allows to prevent operating room (but not atmospheric) pollution by volatile agents. The aim of this study was to create a prototype filter for volatile agents and to test its adsorption efficacy during an ex-vivo simulated conventional cardiopulmonary bypass test.
Methods:
We carried out bench tests to conceive a prototype filter that could prevent room and environment pollution without damaging membrane oxygenators. We performed the tests at the Engineering Center for Circulatory Assistance Laboratory, Dante Pazzanese Institute of Cardiology, São Paulo, Brazil. Bench tests included simulation of integral adsorption tests, filter dimensions and design, flow versus pressure curve, sizing, and tightness.