Background: Oxygenators, as used in cardiopulmonary bypass (CPB) circuits, are components with good air removal properties. However, under some conditions the semipermeable characteristics of hollow fibers allow air to accidentally enter the blood side of the CPB circuit. This may occur when a fluid in motion is stopped suddenly by which the rapid change in momentum may cause a relative negative pressure drop, the so-called hammer effect. The hammer effect is not yet described in literature related to CPB. The aim of this in vitro study was to reproduce the hammer effect.